Blood Pressure Measurement (Manual)

Manual blood pressure uses a cuff and stethoscope to find the pressure at which blood-flow sounds first appear (systolic) and finally disappear (diastolic).

16 steps5 criticalLimit 5 minAlways tested

Steps

  1. 1Perform hand hygiene and identify the patient using two identifiers (for example name and date of birth)
  2. 2Explain the procedure to the patient and ask if they have any questions
  3. 3Critical stepSeat the patient quietly for at least 5 minutes with feet flat on the floor, back supported and legs uncrossed; no talking
  4. 4Check for conditions that rule out an arm (fistula, mastectomy on that side, IV line, clots) and use the other arm if needed
  5. 5Critical stepRemove or rearrange clothing so the cuff and stethoscope are on bare skin; do not leave a rolled-up sleeve under the cuff
  6. 6Select the cuff size from the upper-arm circumference: bladder width about 40 percent of it and bladder length 80 to 100 percent of it
  7. 7Support the arm at heart level (resting on a table or held by you, not held up by the patient)
  8. 8Critical stepPalpate the brachial artery, then center the cuff bladder over it with the lower edge about 1 inch above the bend of the elbow; fit the cuff evenly and snugly
  9. 9Palpate the radial pulse and inflate the cuff until the pulse is no longer felt; note that level, deflate rapidly, and wait 30 seconds
  10. 10Place the stethoscope bell or diaphragm lightly over the brachial artery
  11. 11Critical stepInflate the cuff rapidly to 30 mm Hg above the level where the radial pulse disappeared
  12. 12Deflate slowly at 2 to 3 mm Hg per second; the first Korotkoff sound is the systolic pressure
  13. 13Continue deflating at about 2 mm Hg per second; the point where the sounds disappear completely is the diastolic pressure
  14. 14Deflate the cuff completely and remove it from the patient's arm
  15. 15Tell the patient the reading, clean the stethoscope and cuff, and perform hand hygiene
  16. 16Document the reading, the arm used and the patient's position, and report significant deviations per facility policy

What graders are watching for

Tips
  • Wrong cuff size is the most frequent error: too small reads falsely high, too large reads falsely low
  • Rest first: the patient should avoid caffeine, exercise and smoking for at least 30 minutes beforehand, and neither of you should talk during the rest or the reading
  • An unsupported back, crossed legs, or an arm hanging below heart level can all raise the reading

Common mistakes

  • Wrong cuff size (the AHA statement calls miscuffing the most frequent error), giving a falsely high or low number.
  • Skipping the rest period or letting the patient or you talk during the reading.
  • Skipping the palpated estimate, so the inflation target is a guess.
  • Deflating faster than the 2 to 3 mm Hg per second the sources give.
Practice this skill

Practise the sequence with critical-step highlighting and a rationale for each step.

Start your 7-day free trial

7 days free, then $9.99/month. Cancel anytime.

Sources and how we check them: How we verify.